Closed-loop control logic for mass flow controllers: How to achieve millisecond-level response?

2026-01-29

Word Size

The millisecond-level response capability of a mass flow controller (MFC) stems from its highly integrated and optimized closed-loop control system. At the heart of this system is the goal of enabling the actual gas flow to track the setpoint instantaneously and with pinpoint accuracy. The underlying logic is a continuously running micro-cycle of “measurement—comparison—correction.”

I. The Core Component of Closed-Loop Control Logic
Measurement: The MFC employs a thermal principle (the mainstream method) to directly measure mass flow rate. The sensor converts the heat changes caused by gas flow into electrical signals, which, after amplification and digitization, serve as the system’s current actual flow rate value.
Comparison: The internal microprocessor (MCU) compares the digitized actual flow value with the setpoint value input from the outside in real time, calculating the instantaneous error.
Correction: The MCU sends the error signal to its core control algorithm (typically an optimized PID algorithm or one of its variants). After high-speed computation, the algorithm generates a control signal that actuates a proportional valve—such as a piezoelectric valve or a solenoid valve—precisely adjusting the valve’s opening degree. This, in turn, modifies the gas flow rate, bringing the actual flow closer to the setpoint value.
This “measurement-comparison-correction” cycle is executed continuously at a frequency that can reach the kHz level, thereby forming a dynamic and highly responsive negative feedback closed loop.
II. Key Technologies for Achieving Millisecond-Level Response
To achieve millisecond-level response, MFC has undergone deep optimization at the following three levels:
1. Sensor and Hardware Optimization
Low-thermal-capacity sensing design: By employing capillary or microelectromechanical (MEMS) sensors, the thermal inertia of the thermal system is greatly reduced, enabling instantaneous detection of flow changes.
High-speed control valves—particularly piezoelectric valves—exploit the inverse piezoelectric effect of piezoceramics to achieve mechanical deformation on the microsecond scale, offering extremely rapid response times.
2. Intelligent control algorithm
Adaptive PID and Feedforward Control: Traditional PID parameters are difficult to keep optimal under all operating conditions. The advanced MFC employs an adaptive algorithm that automatically tunes the PID parameters based on the flow rate. At the same time, feedforward control is introduced to predictively actuate the valve immediately upon changes in the setpoint, significantly reducing initial errors and overcoming the delay inherent in pure feedback systems.
3, System Integration and Calibration
Integrated “Sensor-Control-Valve” System: The three components are tightly encapsulated within an adiabatic, compact unit, significantly reducing dead volumes and delays in the gas flow path and ensuring the compactness of the control loop.
Full-scale, high-precision calibration at the factory: For specific gases, the device undergoes full-scale, multi-point data calibration and linearization at the factory, with the resulting data stored in the MCU. This ensures accurate measurements across the entire measurement range and rapid convergence of control performance.
In summary, the millisecond-level response of the MFC is the result of the synergistic interaction among three key factors: its hardware (fast sensors + fast valves), intelligent algorithms (adaptive PID + feedforward), and system integration. By employing a closed-loop control at a specific frequency, the MFC swiftly suppresses flow fluctuations in their nascent stage, thereby achieving precise and rapid control of gas flow.


Related Products

ACU10FA


The ACU10FA analog-type mass flow controller features fast response and high precision. It offers a wide control range, with a flow rate capability spanning from 5 SCCM to 30 SLM, and can accurately measure and regulate flows as low as 1 SCCM—making it ideal for controlling even the tiniest gas flows.

ACU10FD


The ACU10FD digital mass flow controller consists of a mass flow sensor, laminar flow separator, flow controller regulating valve, and amplification control circuit. It features digital signal input and output, offering advantages such as short preheating time, minimal zero drift, and high reliability.

ACU10FDR


The ACU10FDR low-pressure-drop gas mass flow controller is equipped with a proprietary high-flow electromagnetic valve, making it ideal for controlling mass flow even when system pressure is nearly unavailable. Its minimal pressure differential ensures the least possible impact on the system, enabling precise flow control under near-atmospheric conditions while significantly reducing system response time. We can customize the valve and PID control features based on your specific application parameters, guaranteeing fast, stable performance with minimized pressure loss.

ACU10FC


The ACU10FC differential-pressure mass flow controller employs a differential-pressure-based mass measurement principle that features no thermal drift and zero response lag. It can simultaneously display and output parameters such as instantaneous flow rate, totalized flow, pressure, and temperature. When control functionality is required, an electromagnetic proportional control valve can be added, along with built-in PID regulation capabilities. Unlike thermal mass flow meters, the ACU10FC eliminates any additional "specific heat capacity" error when using gas conversion factors. As a result, even if the gas being measured differs from the calibration gas used at the factory, the measurement accuracy remains unaffected.

ACU10P Series


The ACU10P series pressure controllers/pressure gauges feature a straight-through design and cover a measurement range from 100 kPa to 10 MPa (both absolute and gauge pressures are supported). In addition, differential pressure transmitters are also available.